Appendix I: Path Reduction Module
96
LISST-
200X User’s Manual
Appendix I: Path Reduction Module
Introduction
Transmission
Limit
Increasing
Limits by
Reducing Path
Concentration
Limits
The LISST-200X can measure over a wide range of particle
concentrations. However, as concentration increases, an increasing
fraction of detected light is scattered more than once before reaching
the detector. This multiple scattering can distort the apparent size
distribution, specifically by skewing it toward smaller sizes. The Path
Reduction Module (PRM) reduces this effect, extending the operating
range to higher concentrations.
There is no abrupt threshold when multiple scattering suddenly
becomes excessive. Rather, it is a gradual effect: as overall
scattering increases, the proportion of multiple scattering increases
as well. The effect is also highly dependent on particle size. The
relationship is complex. However, to assess the risk of multiple
scattering it is generally sufficient to consider only the optical
transmission. When transmission is below 30%, multiple scattering is
likely to have a noticeable effect on size distribution data
, “inventing”
fine particles. Transmissions as low as 10% may yield usable data,
but would require special care in processing and evaluating data.
If transmission is too low, meaning multiple scattering is too high, for
accurate measurements, the optical path that the laser passes
through must be reduced. This reduces the number of scattering
interactions between the laser beam and particles, thereby reducing
the proportion of multiple scattering.
Transmission and maximum concentration vary inversely with path
length. Therefore, reducing the path by a factor of 2 increases the
measurable concentration by a factor of 2. Sequoia’s Path Reduction
Module (PRM) for the LISST-200X reduces the path from 25 mm to
5 mm, for a factor of 5 increase in maximum concentration.
The table below shows the concentration limits for various particle
sizes, with and without the PRM.
Mean particle
diameter
Wentworth
grades
LISST-200X Estimated Maximum concentration in mg/L
µm
Phi
25mm Path (Standard)
5mm Path (with PRM installed)
1.95
9.00
Clay
83
416
3.9
8.00
Very Fine Silt
166
831
7.8
7.00
Fine Silt
332
1662
15.6
6.00
Medium Silt
665
3324
31.25
5.00
Coarse Silt
1332
6659
62.5
4.00
Very fine sand
2728
13638
125
3.00
Fine sand
5327
26637
250
2.00
Medium sand
10655
53273
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...