LISST-
200X User’s Manual
61
LISST-200X Operation Details
E.
LISST-200X Command Details
NOTE: Command are shown in upper case for clarity, but are not case-sensitive.
1.
Display Commands
CONFIG
Display instrument hardware configuration
Syntax:
CONFIG
Description:
The current hardware configuration including Auxiliary I/O connector
configuration
Example:
L200x:> CONFIG
Serial Number: 2001
Firmware Version: 1100 (1.100)
VCC: 59371
Full Path Length: 2500 (25 mm)
Eff. Path Length: 2500 (25 mm)
BioBlock Installed: 0
Start Condition: Digital input (4)
Start Condition Data: 0
Stop Condition: Switch (3)
Stop Condition Data: 0
Measurements Per Average: 10
Sample Interval: 1 seconds
Burst Mode: 0
Samples Per Burst: 5
Burst Interval: 25 minutes
Transmit Raw Data: 0
Lifetime Sample Count: 529554
Lifetime Laser On Time: 1703980 seconds
Endcap Configuration: 0 (Analog & digital in)
Wiring of MCBH6MP connector:
1: Ground
2: Digital input 1
3: Digital input 2
4: +12V output
5: Analog input 1
6: Ground
DD, DIR
Display File Directory
Syntax:
DD or DIR
Description:
Display current disk directory in DOS type format. Includes total bytes
used and bytes available.
Example:
L200x:> DD
. <DIR> 2099-03-14 03:35:48
.. <DIR> 2099-03-14 03:35:48
L2390658.RBN 1,612,440 2019-08-29 14:55:28
L2411501.RBN 3,253,800 2019-09-03 07:58:20
L2480807.RBN 736,800 2019-09-06 09:40:14
3 file(s) 7,608,360 bytes
2 dir(s) 1,016,381,440 bytes free
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...