LISST-
200X User’s Manual
99
Appendix I: Path Reduction Module
2. Software Configuration with PRM
Before
deploying the instrument with the PRM, you must indicate to the instrument’s
onboard firmware that the PRM is installed. Then, data files produced by the instrument
will properly include that information, and the software will properly account for the
reduced path.
1
Open the LISST-200X software and connect your instrument. Press the
Configure Instrument button to see this screen:
2
Select the Pathlength Reduction Module option. Note that the PRM cannot be
used with the BioBlock (described in Appendix H: BioBlock Anti-fouling Wiper).
3
Press the Upload to Instrument button, to make the setting permanent.
You can confirm the setting at any time by clicking on the blue tab at the bottom
of the screen to open the configuration panel:
After installing the PRM, you MUST make a background measurement.