LISST-200X Operation Details
66
LISST-
200X User’s Manual
Cautions:
Be very careful
with this setting!
Issue a DS command to verify the status of the store mode setting. A
warning will be displayed as part of the DS status if the store mode is
disabled:
SAVEPSD
Control saving of processed PSD files (firmware 2.1 or higher)
Syntax:
SAVEPSD x
Description:
If x = 1, during sampling the instrument will generate a .CSV file
containing processed particle size distributions, in addition to the normal
raw .RBN file.
If x = 0, only the normal .RBN files will be saved.
For more details, see Autonomous Real-time Data Processing on page
52. This command requires firmware version 2.1 or higher.
Example:
L200x:> SAVEPSD
PSD data will be saved during logging.
SB
Set Samples per Burst
Syntax:
SB x, where x is the number of samples per burst.
Each sample is taken at the sample interval (in seconds) set by the SI
command.
If no value follows command, prompts will be displayed for the value.
Example:
L200x:> SB 10
New Samples per Burst: 10
SC
Set Clock with time and date
Syntax:
SC yyyy/mm/dd hh:mm:ss
Where yyyy=year, mm=month, dd=day, hh=hour (24 hour format),
mm=minute, ss=seconds
If no values follow the “
SC
” or “
sc
” command, prompts for entering the
time and date will be displayed.
Example:
L200x:> SC 2020/12/20 13:20:18
Time set to 2020/12/20 13:20:18
SENDDATA
Transmit data during sampling (firmware 2.1 or higher)
Syntax:
SENDDATA x
Description:
The value of x sets
whether data will be transmitted from the instrument’s
RS232 interface during sampling.
Only options 0 and 4 are suitable for
normal operation.
0 = Do not send data (default)
1 = Multi-line troubleshooting data
2 = NMEA-format brief data (for glider applications)
3 = Single-line raw (CANNOT be processed by LISST-200X software)
4 = Processed PSD data; activates real-time processing and sends
processed data after each sample. For details, see Autonomous Real-
time Data Processing on page 52. This option requires firmware version
2.1 or higher.
Example:
L200x:> SENDDATA 0
Setting: 0 (no data)
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...