LISST-
200X User’s Manual
35
LISST-200X Operation Details
7
After configuring the instrument settings, they must be uploaded
onto the instrument. A message will be displayed on the top of
the screen when you have unsaved changes to the instrument
configuration
Click on the ‘Upload to Instrument’ button to load the
configuration onto the LISST-200X. A confirmation message will
be displayed when the upload is finished.
Upload
Configuration
8
•
The instrument is now configured for deployment,
however,
the sampling program is not yet running
.
•
You MUST run the sampling program in order to collect
data. But even when running, the sampling program will wait
for the programmed start condition before collecting data.
See Start and Stop Conditions on page 57, and the steps
above, for details about start conditions.
•
When the sampling program is running, you will receive a
confirmation message that the program has started, and the
instrument will become disconnected from the computer. At
this point the instrument is ready to be deployed.
•
The green LED on the connector endcap will indicate the
instrument status:
- If waiting for the start condition, every few seconds it will
blink twice quickly.
- If logging, it will light while collecting each sample.
Run Sampling
Program
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...