LISST-
200X User’s Manual
65
LISST-200X Operation Details
PD,
STOPDATA
Set Stop Condition Data
Syntax:
PD x or STOPDATA x where x is the stop condition data as described
below
Description:
The PD command sets the stop condition data to be used when the
collection data. The stop condition data is used with the Stop Condition
settings as follows.
•
If the Stop Condition is Depth Stop (option 1) the input will be stop
depth in meters.
•
If the stop condition is set to Time/Date Stop (option 2) the input for PD
will be the stop date and time.
•
If the Stop Condition is Fixed Number Stop (option 5) the input will be
the number of samples to collect before stopping.
•
The PD setting is ignored for Mechanical Switch Stop (option 3),
Digital Input Stop (option 4) or Maximum memory or Low Battery Stop
(option 6).
Example:
If Stop Condition =1 (Depth Stop):
L200x:> PD 3
Depth stop at 3 meters
If Stop Condition = 2 (Time/Date Stop):
L200x:> PD 2016/12/31 23:59:59
Time stop at 2016/12/31 23:59:59
If Stop Condition = 5 (Fixed Number of Samples):
L200x:> PD 100
Stop after 100 samples.
PSDMODEL
Set particle model used for on-board size distribution calculations
(firmware 2.1 or higher)
Syntax:
PSDMODEL x
Description:
If x = 0, PSD processing with use the spherical particle model.
If x = 1, the irregular (also called “random”) particle model will apply.
This setting applies only when real-time PSD processing is active. It is
available only in firmware 2.1 and higher. Also see SAVEPSD,
SENDDATA, and Autonomous Real-time Data Processing on page 52.
Example:
L200x:> PSDMODEL 1
Optical Model: 1 (irregular)
SAVEDATA
Save Data Setting
—USE WITH CAUTION
Syntax:
SAVEDATA x, where x is 1 or 0
Description:
SAVEDATA 0 turns off storing of raw data in the instrument’s memory.
This is only for unusual circumstances.
Example:
L200x:> SAVEDATA 1
1: Raw data will be saved during logging.
L200x:> SAVEDATA 0
WARNING: Data will NOT be saved during logging!
Is this really what you want [N] ? n
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...