LISST-
200X User’s Manual
31
LISST-200X Operation Details
•
Multiple cleanings or using better filtered water may be
required to get the best possible background.
•
When you have an acceptable background press the
Save button to store the background on the
instrument. You also have the option to save the
background to a file on your computer. After the
background is saved, select the ‘Save as .BGT File’
button. The LISST-200X handles the background files
differently than the LISST-100X. The background file
is stored on the instrument and is saved as part of
every data file that is recorded. When processing a
data file the background will be automatically
extracted and used during processing. Saving,
tracking and selecting a background file to use for
processing is no longer required. The background
stored on the instrument will continued to be saved in
new data files until a new background is recorded.
Background Accepted
and saved.
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...