LISST-
200X User’s Manual
49
LISST-200X Operation Details
6
•
If your transmission values are < 0.10 (or 10%), the
water is too turbid. Disregard these data.
Data with transmission
values < 10% should
be disregarded.
7
•
If your transmission values are > 0.995 (or 99.5%),
the water is too clear. Disregard these data.
Data with transmission
values > 99.5% should
be disregarded.
8
Summarizing
•
Disregard data if transmission is > 0.995 (> 99.5%).
•
Disregard data if transmission is < 0.10 (< 10%).
•
Be wary of data collected at transmission values
between 0.98 and 0.995
– low signal-to-noise ratio.
•
Be wary of data collected at transmission values
between 0.30 and 0.10
– generally decreasing data
quality as the transmission decreases below 0.30
(30%).