Appendix H: BioBlock Anti-fouling Wiper
Appendix H: BioBlock Anti-fouling Wiper
The BioBlock is an optional accessory that periodically wipes the windows of the LISST-
200X, drastically reducing biological growth (biofouling) on the windows. Biofouling on the
windows creates excess background scattering, which can degrade the data from any long-
term deployment in biologically active waters.
Hardware Requirements and Compatibility
All LISST-200Xs contain electronics for driving a BioBlock, via a 3-pin connector on the
endcap (see page 86 for connector details). The hardware and fittings required for mounting
to the LISST-200X are supplied with the BioBlock.
Due to mechanical constraints, the BioBlock cannot be installed on instruments in the
LISST-Black/HAB configuration (described on page 90), nor with a path length reduction
module.
The BioBlock can be installed on a LISST-200X mounted to the optional welded cage
supplied by Sequoia Scientific.
Power Requirement
During the brief period (less than 2 seconds) while it is opening or closing, the BioBlock
roughly doubles the power consumption of the LISST-200X. This is compatible with the
large battery housing (see page 39), but not the small battery housing (page 36), or the
USB-powered cable. The USB-powered cable can be used for communication when the
BioBlock is connected, but you must connect the large battery housing or another power
source while operating it.
Firmware Requirement
If connecting a BioBlock for the first time, check that your LISST-200X has an appropriate
firmware version. In the LISST-200X software, connect to the instrument, then click on the
blue tab at the bottom of the main window to open the instrument information pane, which
shows the firmware version:
If the firmware version starts with the digit 1, it must be at least 1.461 to support the
BioBlock. If it starts with 2, it must be at least 2.020.
BioBlock Activation
To operate automatically, the BioBlock must be activated with a software command. In the
LISST-
200X software, connect to the instrument and click “Configure Instrument”. Then click
“Accessories” to open the panel shown below, and select the BioBlock option.
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...