Appendix H: BioBlock Anti-fouling Wiper
94
LISST-
200X User’s Manual
You can also activate the BioBlock in the terminal window, by typing the command BB
AUTO in the terminal window, or deactivate it with BB OFF. The commands BB OPEN and
BB CLOSE in the terminal will immediately open or close it.
When the BioBlock is activated, it will automatically open (wiping the windows) whenever
the sampling program (set up as described starting on page 32) is running. It will close
whenever the sampling program pauses for at least 50 seconds.
NOTE: your sampling
program must either use burst mode, or a long sample interval, to schedule periodic
gaps of at least 50 seconds when the BioBlock will wipe and close. You must test the
sampling program to verify the BioBlock schedule behaves as you expect.
Mechanical Installation
Required Tools:
1/16-inch hex L-key
9/64-inch hex L-key
1. Put bracket A onto the end of the 200X. The tab on the front edge will set the
position.
2. Using
9/64” L-key, remove the screw on bracket B. Slide the bracket over the
connector end of the 200X. Place it approximately 4 inches/10 cm away from the
large bracket.