LISST-
200X User’s Manual
67
LISST-200X Operation Details
SENSORPOWER
Control 12V power to auxiliary connector during logging
Syntax:
SENSORPOWER x, where x is 1 or 0
Description:
If x = 1, the 12V power output on the auxiliary connector (if it is
suitably configured) will turn on during data collection with the GO
command. If running in burst mode, the power will turn off between
bursts. Logging will also incorporate a warmup time, set with the
SENSORWARM command.
If SENSORPOWER is entered without an x argument, it will prompt
for a value and for a sensor warmup time.
Example:
L200x:> SENSORPOWER 1
Power output will turn on during logging, with 5 second
warmup
SENSORWARM
Warmup time when powering external sensors
Syntax:
SENSORWARM x, where x is in seconds
Description:
If SENSORPOWER is on, SENSORWARM sets the number of
seconds the LISST-200X will wait for external sensors to warm up
before it proceeds with sampling.
Example:
L200x:> SENSORWARM 10
Power output, when used, will turn on with 10 second
warmup.
Power output during logging is on.
SI
Set Sample Interval
Syntax:
SI x, where x is the number of seconds between samples, from 1 to
10,000.
Description:
In either fixed-rate or burst mode, the sample interval is the number of
seconds between two consecutive samples, each composed as an
average of a number of measurements (specified by the MA command).
Example:
L200x:> SI 5
Seconds between samples: 5
SP,
STOPCOND
Set Stop Condition
Syntax:
SP x or STOPDATA x, where x is the stop condition code described below
Description:
The SP command sets the stop condition to be used when collecting data.
The stop condition options are:
1 = Depth Stop
2 = Time/Date Stop
3 = Mechanical Switch Stop
4 = Digital Input Stop
5 = Number of Samples Stop
6 = Maximum memory or Low Battery Stop
If no value follows command, prompts will be displayed for the value. After
setting a stop condition of Depth, Time/Date or Number of Samples, use
STOPDATA to enter an appropriate parameter value.
Example:
L200x:> SP 5
Stop condition: 5
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...