LISST-200X Operation Details
62
LISST-
200X User’s Manual
DEPTH
Display current depth
Syntax:
DEPTH
Description:
The current depth is displayed to the screen.
Example:
L200x:> depth
0.000 m (680 counts)
DS
Display current status information
Syntax:
DS
Description:
The instrument settings and status are displayed to the screen. The
format may change with different firmware and hardware versions.
Example:
L200x:> DS
LISST-200X Status and Settings
Serial number: 2021
Endcap configuration: 2 (Dual analog out)
Firmware Version: 1.470 Feb 8 2021 21:06:43
Current Date/Time: 2019/01/09 22:12:50
Current Day of the Year: 009
Supply voltage: 14.74 V
Operating mode: Fixed rate mode
AutoStart: OFF.
Delay Start with 0 minute delay
Stop on full memory or low battery
Measurements per Average: 32
Ambient light subtraction: on.
Sample Interval: 1
Path length: 25 mm
Standard status messages will be sent during logging.
Data will be saved in files of unlimited size.
AUV mode: off
Free memory: 1016 Mbytes
DS 1
Display current status information, compact form
Syntax:
DS 1
Description:
Instrument status displayed in simplified format for reading by software
Example:
L200x:> ds 1
LISST-200X Current Status and Settings
SN = 0
OM = 0
ST = 5
TD = 0
SP = 6
PD = 0
MA = 32
SI = 1
BI = 1
SB = 1
AS = 0
BIOBLOCK = 0
STUBE = 0
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...