LISST-200X Operation Details
60
LISST-
200X User’s Manual
D.
LISST-200X Command Summary
Display Commands
CONFIG
Display instrument hardware configuration information
DD, DIR
Display current disk directory
DEPTH
Display current depth
DS, DS 1
Display current instrument status information
DT
Display current time and date
HELP, ?
Display general help messages and command list
SW
Display mechanical switch position
TEMP
Display current temperature
VIN
Display input supply voltage
Setup Commands
AS x
Set autostart
BI x
Set burst interval
MA x
Set samples per average equal to x
MODE, OM x
Set operating mode (burst or fixed-rate)
PD x, STOPDATA x
Set stop condition Data
PSDMODEL x
Set model used for PSD processing (if active)
SAVEPSD x
Set whether PSD data will be saved in on-board memory
SI x
Set sample interval
SP x, STOPCOND x
Set stop condition
ST x, STARTCOND x
Set start condition.
SB x
Set samples per burst
SC yyyy/mm/dd hh:mm:ss Set clock with time and date
SENDDATA x
Control real-time transmission of data via RS232
SENSORPOWER x t
Set whether power output turns on during logging
SENSORWARM t
Set external sensor power warmup time
TD x, STARTDATA x
Set start condition data
ZD
Reset depth sensor offset
Acquisition/Action Commands
DL filename
Delete file
GO
Start data collection using current parameters
GX
Grab sample and transmit it
XR
Transmit raw data when average saved
ZS
Acquire 20 samples and evaluate them for quality.
ZZ
Go into deep sleep mode (minimum power consumption)
Содержание LISST-200X
Страница 2: ...Intentionally blank...
Страница 3: ...LISST 200X User s Manual Page iii This document is copyrighted by Sequoia Scientific Inc...
Страница 4: ...Intentionally blank...
Страница 6: ...Intentionally blank...
Страница 8: ...Page viii LISST 200X User s Manual Intentionally blank...
Страница 85: ...LISST 200X User s Manual 77 Appendix A Technical Specifications...