Appendix I: Path Reduction Module
102
LISST-
200X User’s Manual
quartz. Clean any deposits or growth from the outside of the cylinder to ease
removal of the ring.
•
The o-ring, item 4, is not used as a seal, but as a spring to keep the cylinder
pressed against the small window. Remove the o-ring to inspect it. Replace
it if it is brittle, cracked or sticky.
3
•
Install the black plastic retaining ring onto the quartz cylinder. There should
be about a 3mm gap between the cylinder and the plate, as shown below.
The cylinder should be able to move freely.
•
Reattach the PRM Plugs using the two screws to keep all parts together.
•
Wrap the optics with soft tissue and store the assembly in a safe place.