LISST-200X Operation Details
70
LISST-
200X User’s Manual
ZS
Collect and transmit background scattering data
Syntax:
ZS
Description:
Acquires 20 averaged sample and displays the result to the screen.
Evaluates the data for quality as a background measurement.
Example:
Input: ZS
Output: {
…. 36 ring 24 Aux parameters
}
repeat 20 times, then messages about the quality of the data.
Notes:
The ZS command does not store the sample to a datafile.
ZZ
Go into deep sleep mode (minimum power consumption)
Syntax:
ZZ
Description:
Sends LISST-200X in to low power sleep mode. Instrument will send a
message every 30 seconds to indicate it is sleeping. It can be wakened at
any time by sending 2 control-C characters about ½ second apart.
Example:
L200x:> ZZ
Low-power sleep.
zzz...Press Stop or <Ctrl>C to wake up.
zzz...Press Stop or <Ctrl>C to wake up.