LISST-100X User’s Guide
119
Step 5)
Use large tie-wraps to secure the
BioBlock body to the LISST-100X, as
shown in the photo on the right. The tie-
wraps should be placed over the bolts
extending from the mounting brackets.
Trim excess ends of tie-wraps.
Step 6)
Connect the BioBlock cable to the
LISST-100X and configure the instrument
using the instructions on the following
pages.
Then verify normal operation by cycling
the system using the LISST.EXE software.
The BioBlock shutter should position itself
smoothly during cycling, with no binding.
If necessary, loosen the mounting screw
slightly and adjust the position of the
BioBlock slightly to improve operation, and
then re-tighten the screw.
The figure on the left shows the fully
assembled BioBlock on the LISST-100X.