84
LISST-100X User’s Guide
Syntax: SP x or sp x, where x is the stop condition code described below
Description: The SP command sets the stop condition to be used when collecting
data. The stop condition options are:
1 = Depth Stop
2 = Time/Date Stop
3 = Mechanical Switch Stop
4 = Digital Input Stop
5 = Fixed Number of Samples Stop
6 = Maximum memory or Low Battery Stop
If no value follows command, prompts will be displayed for the value.
Example: input: SP 5
output: New Stop Condition Setting: Fixed Number Stop
PD
Set Stop Condition Data
Syntax: PD x or pd x, where x is the stop condition data as described below
Description: The PD command sets the stop condition data to be used when the
collection data. The stop condition data is used with the Stop
Condition settings as follows.
If the Stop Condition is Depth Stop (option 1) the input will be
stop depth in meters.
If the stop condition is set to Time/Date Stop (option 2) the
input for TD will be the stop date and time.
If the Stop Condition is Fixed Number Stop (option 5) the input
will be the number of samples to collect before stopping.
The TD setting is ignored for Mechanical Switch Stop (option
3), Digital Input Stop (option 4) or Maximum memory or Low
Battery Stop (option 6).
Example: input: PD 3 (if Stop Condition =1 (Depth Stop))
output: New Stop Condition data = 3
Stop Condition: Depth Stop at 3 meters
input: PD12/31/05 23:59:59 (if Stop Condition = 2 (Time/Date Stop))
output: New Stop Condition data = 12/31/05 23:59:59
Stop Condition: Time Stop at 12/31/05 23:59:59
input: PD 2 (if Stop Condition = 5 (Fixed Number of Samples Stop))
output: New Stop Condition data = 2
Stop Condition: Fixed number Stop at 2 samples
SI
Set Sample Interval
Syntax: SI x or si x, where x is the number of seconds between samples, from
1 to 10,000.
Description: In Fixed Rate Mode (OM = 3) or Burst Rate Mode (OM = 2), the
sample interval is the number of seconds between two consecutive
samples, each composed as an average of a number of
measurements (specified by the MA command).
Example: input: SI 5
output: New Seconds between Samples: 5
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...