LISST-100X User’s Guide
19
measurements are averaged into a sample, and it is this sample
average that is stored at a maximum sample rate of 1 Hz. For
the example shown above, the instrument is set to sample in the
Fixed Sample Rate mode at a 1 Hz rate with 10 measurements
per average. This average is obtained in 0.34 seconds.
Step 13: Setting
Start Conditions
After selecting the Operating mode the start and stop conditions
can be selected. Choose the
Start Condition
tab at the top of the
window.
There are five options: Depth, Time, External Mechanical Switch,
External Digital Input, and Time Delay. Select the mode by
clicking on the button next to its label. Select the correct
parameters as required. For this example let’s select the
External Mechanical Switch Start Condition.
Step 14: Setting
Stop Conditions
Similarly the Stop conditions can also be selected. Click on the
Stop Condition
Tab to open the Stop Condition window. The
available stop conditions are: Depth, Time. External Mechanical
Switch, External Digital Input, Fixed number of Samples, and
Maximum Memory/Low Battery.
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...