LISST-100X User’s Guide
6
Choose the Stop Conditions Tab to select the conditions
when sampling should stop.
The available stop conditions are: Depth, Time. External
Mechanical Switch, External Digital Input, Fixed number of
samples, and Maximum Memory or Low Battery.
If the Stop Condition is Depth, External Mechanical Switch,
or External Digital Input and the Start Condition is also one
of these three options the Base Program on the instrument
will return to waiting for the Start condition. For example, if
the start and stop conditions were set to External
Mechanical Switch then multiple sampling sessions can be
obtained by flipping the switch on and off. This is useful
when doing profiles. It eliminates the need to communicate
with the instrument between profiles.
.
7
Select
Apply
or
OK
buttons to configure the instrument with
the current settings. If the
Apply
button is pressed the
program will return to the current window. Returning to the
Instrument Status
window will display a summary of the
current settings.