LISST-100X User’s Guide
81
LISST-100X Command Details
Display Commands
DB
Display Battery Voltage
Syntax: DB or db
Description: The current battery voltage is displayed to the screen.
Example: input: DB
output: Current Battery voltage: 8.90 volts
DD
Display Disk Directory
Syntax: DD or dd
Description: Display current disk directory in DOS type format. Includes total
bytes used and bytes available.
Example: input: DD
results: L100x:>dd
LISST-100X Disk Directory
Volume in drive C is NONAME
Volume Serial Number is 778B-155F
Directory of C:\
L040305_.BIN 4,136 03-05-04 6:30p
L183705.BIN 1,672 03-05-04 6:37p
2 file(s) 5,808 bytes
0 dir(s) 15,933,440
bytes free
L100x:>
DS
Display current status information
Syntax: DS or ds
Description: The instrument settings and status are displayed to the screen.
Example: input: DS
output:
LISST-100X Current Status and Settings
Serial number = 1004
Operating Mode: Burst Mode
Start Condition: Time Start at 03/10/04 09:22:00
Stop Condition: Fixed Number Stop at 1 samples
Measurements per Average: 10
Sample Interval: 1
Burst Interval: 1
Samples per Burst: 0
Battery Voltage is 8.90
Current Date/Time: Wednesday, March 10, 2004 10:18:49
input: DS 1
output:
LISST-100X Current Status and Settings
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...