LISST-100X User’s Guide
69
STEP BY STEP PROCEDURE: CONFIGURING DATA COLLECTION
Step
Action
Result
1
Connect the instrument to the PC and start the LISST-SOP
program.
LISST-SOP
open and
communicating
with LISST-
100X.
2
Open the Operating Modes window by choosing
Operating
Modes
from the
LISST
menu or by pressing the
button
on the toolbar.
When first opening the Operating Modes window the LISST-
SOP program will send a Ctrl-C to the LISST-100X. If a
program is running it will be stopped. This is required so that
the LISST-SOP program can get the status information from the
LISST-100X.
To set the LISST-100X data logger clock to the computer clock
press the
Set Clock
button. Press the
Query Instrument
button
to refresh the display.
Instrument
Status
Displayed.
Clock set