88
LISST-100X User’s Guide
Example: Input: DL L159*.dat
Output: Are you sure (Y/N)…
Cautions: WARNING: Make sure that the file being delete has already been
offloaded before deleting the file. Once the file is delete it can not be
recovered.
GO
Start Data Collection using current Settings
Syntax: GO or go
Description: Starts Fixed Rate or Burst Mode Data collection using current
settings.
Example: Input:
GO
Output: Waiting for start conditions…
Cautions: To stop data acquisition before it is complete press the Stop button or
CTRL-C
.
GX
Grab sample and transmit it
Syntax: GX or gx
Description: Acquires single averaged sample and displays the result to the
screen.
Example: Input:
GX
Output: {
…. 32 ring 8 Aux parameters
}
Notes: The GX command does not store the sample to a datafile.
XR
Display Raw data to screen when average saved
Syntax: XR x or xr x, where x =1 (yes) or x = 0 (no)
Description: If display of raw data is enabled, the RAW data will be printed to the
serial
port between curly brackets in 40 rows:
{
…. 32 ring 8 Aux parameters
}
Example: input: XR 1
output: Raw data WILL be transmitted when data is stored
Cautions: None
YS
YMODEM Offload
Syntax: YS
filename
or ys
filename
where
filename
is the name of the file to
be offloaded.
Description: Offloads file from Compact Flash Module to PC using YMODEM
offload at 115 Kbaud
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...