If there is a problem with the instrument or if the water or
windows are not clean, error messages will be displayed. Dirty
water or windows will generally cause higher values across the
middle rings. Large bubbles or particles in the water can cause
higher values on the inner rings or left hand side of the display.
High values on the inner rings combined with a lower Laser
power value can also be an indication of optical misalignment.
Step 11:
Configuring
Instrument for
Deployment
The Operating Modes window is used to set all of the
deployment parameters. To open the window, choose
Operating
Modes
from the
LISST
menu or press the
button on the tool
bar. A window similar to the one shown will appear. The window
has four tabs: Instrument Status, Operating Mode, Start
Condition, and Stop Condition.
The Instrument Status page gives information about the
instrument. It also allows previously saved instrument
configuration files (LOP’s) to be loaded on to the instrument.