46
LISST-100X User’s Guide
STEP BY STEP PROCEDURE: BATCH PROCESSING MULTIPLE RAW
DATA FILES
Step
Action
Result
1
Start the LISST-SOP program
Program started.
2
Select Settings from the File menu and click on the
Output Tab. The window below will appear prompting
for the desired output to use for the future data
processing. This selection must be made BEFORE
starting the processing command.
NOTE:
A binary PSD file is always created when
processing raw data. This is used to display the
processed data to the screen.
The ASC type is a space delimited ASCII file
containing all the processed data.
The LOG file is a space delimited ASCII file containing
the raw data from the LISST-100X data logger.
Select file types to be
created.
3
You must also select the inversion model to be used
with the real time data processing: Spherical,
randomly shaped or both.
Select inversion
method(s) to be used
– spherical, randomly
shaped or both..
The Spherical particle model performs the
mathematical inversion under the assumption that the
particles that scattered light are all spheres. Sequoia
employs the so-called full Mie solution when using this
approach. The Mie solution is a generalized solution to
the scattering of light from spheres and is being used
as the standard inversion method by all laser
manufacturers.
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...