82
LISST-100X User’s Guide
SN = 1004
OM = 2
ST = 2
TD = 03/10/04 09:22:00
SP = 5
PD = 1
MA = 10
SI = 1
BI = 1
SB = 0
BB = 0
CT = 0
IC = 0
Current Time = 03/10/04 10:18:49
Battery = 890
Switch = 0
Memory= 15005640
HE
Display general help messages and command list
Syntax: HE or he
Description: Displays the list of command to the screen.
Example: input: HE
output: LISST-100X Commands (followed by 30 lines of text)
Setup Commands
OM
Set Operating Mode
Syntax: OM x or om x
Description: Sets the Operating Mode to one of the following types:
1 = Real Time Mode
2 = Burst Mode
3 = Fixed Rate Mode
OM command only without a parameter will initiate a prompt for the
Operating mode.
Example: input:
OM
3
output: New Operating Mode: Fixed Rate Mode
MA
Set measurements per average
Syntax: MA x or ma x
Where x = number of samples per average
Description: Each recorded or displayed measurement is based on an average of
measurements. The number of measurements per average is set
using the SA command.
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...