LISST-100X User’s Guide
Step 3)
However, before installing the
BioBlock we must first remove one of the
screws on the endcap.
Rotate the LISST-100X to match the
photo. The two endcap screws should be
above the Crossbar Standoff on the left.
One is threaded into the endcap and the
other is longer and threaded into the
instrument housing.
Remove the shorter screw such that the
threaded screw hole is empty. This is the
most clockwise screw of the pair. This
threaded hole will be used to hold the
BioBlock in place.
Step 4)
Slowly slide the BioBlock shutter
assembly straight down between the
Crossbar Standoffs and the windows (the
LISST-100X sample volume), taking care
not to drag the copper plates across the
windows.
The assembly is designed to fit snugly
between the Crossbar Standoffs. If the fit
is too tight the plastic sides of the
assembly may be lightly sanded until the
parts fit together easily.
Holding the BioBlock body firmly against
the LISST-100X, and ensuring that the
shutter assembly is centered in the LISST-
100X sample volume, install and tighten
the mounting screw. This screw is
provided with the unit and is not the screw
that was removed in Step 3.
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...