LISST-100X User’s Guide
115
Step 3)
Carefully insert the Chamber into the
optical path of the LISST-100X. Be careful not to
let the O-rings come into contact with the optical
windows of the LISST-100X.
Step 4)
By reaching in from the top of the chamber
push the Slider out until the chamber is up against
the optics endcap. The Slider should mate to the
cone shape of the transmit optics.
The O-ring in the slider should just be starting to
engage with the wall of the chamber. It may be
necessary to press on the top, bottom and sides of
the Slider to get it to slide out without binding.
Step 5)
Once the slider is positioned such that the
O-ring is engaging the Spacer can be inserted
between the outside wall of the chamber and the lip
of the Slider. The tapered side of the spacer
should go against the Slider lip. This will cause
compression of the O-rings and seal the chamber.
The chamber can now be filled with clean water. It
is highly recommended to do a background
scatterfile after assembly to make sure that the
windows are still clean. It is NOT necessary to
have water around the chamber; only IN the
chamber.
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...