38
LISST-100X User’s Guide
4
A Status bar will be displayed for each file
offloaded. Text in the lower left corner will
display the current file being offloaded.
Offloading files from
memory to the PC
5
The standard offload baud rate is 115K. If you
are using a long offload cable (e.g. 20, 30 or 50
m cable, see Appendix I: LISST Accessories on
page 126) this value may be too high and cause
errors during offloading. If an offload error
occurs, you can change the offload baud rate as
follows:
6
Go to Settings in the File Menu and click the
Serial Port tab. In the File Offload Baud Rate
drop down menu, choose a lower baud rate and
try again.
NOTE: The change in download baud rate is
only supported by LISST-100X firmware
versions higher than 1.900. If you do not see
the dropdown box, your firmware version does
not support change in baud rate. Contact
Sequoia (
) in order to get
a later firmware version that will support
changes in offload baud rates.