60
LISST-100X User’s Guide
STEP BY STEP PROCEDURE: SIMPLE REAL-TIME DATA PROCESSING
Step
Action
Result
1
Connect the instrument to the PC and start the LISST-
SOP program.
Program started.
2
Open the Operating Modes Window and choose the Real-
Time Operating Mode.
Instrument
configured for
Real-Time data
collection.
3
Before starting a Real-Time session the output file options
must be set. This must be done before selecting
Open
Real-Time Session
. Select Output from the Options
menu. The window below will appear.
Output file type
selected.
A binary PSD file must be selected before any other file
type will be stored. This is used to display the processed
data to the screen.
The ASC type is a spaced delimited file containing all the
processed data.
The LOG file is a space delimited file containing the raw
data from the datalogger.
It is strongly recommended that
you always build a LOG file, otherwise troubleshooting the
data will be very difficult if any problems or questions
about the data quality should appear.
You must also select the inversion model to be used with
the real time data processing: Spherical, randomly
shaped or both. For details on these selections, see page
40.
4
Select Open Real-Time Session from the File menu or
press on the
button.
Open Instrument
selected.
Содержание LISST-100X
Страница 2: ......
Страница 3: ...This document is copyrighted by SEQUOIA SCIENTFIC INC...
Страница 4: ......
Страница 5: ......
Страница 7: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 19: ...LISST 100X User s Guide 7...
Страница 20: ...8 LISST 100X User s Guide...
Страница 61: ...LISST 100X User s Guide 49...
Страница 65: ...LISST 100X User s Guide 53 Notes...
Страница 71: ...LISST 100X User s Guide 59 Notes...
Страница 92: ...80 LISST 100X User s Guide consumption...
Страница 110: ...98 LISST 100X User s Guide...