Remote control commands
R&S
®
NRX
389
User Manual 1178.5566.02 ─ 09
Parameters:
<auto>
ON | OFF | ONCE | LFR | UFR
CALibration<Sensor>:ZERO:FAST:AUTO
[<auto>]
Effective for trace measurements.
Performs fast zeroing. Since the commands are processed very quickly, they are not
overlapping.
The command is not fully R&S
NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S
NRP2 emulation using
"NRP2"
.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<auto>
ON | OFF | ONCE | LFR | UFR
See
CALibration<undef>:ALL:ZERO:FAST:AUTO
[<auto>]
Applies to all connected power sensors. See
CALibration<Sensor>:ZERO:FAST:
Suffix:
<undef>
.
1 to n
No suffix required.
Parameters:
<auto>
ON | OFF | ONCE | LFR | UFR
13.12
Running selftests
Used for testing the connected power sensors and the R&S NRX.
Further information:
●
.......................................................................................... 390
CALibration<Sensor>:TEST:DEViation?
........................................................................... 390
CALibration<Sensor>:TEST:REFerence?
......................................................................... 390
...............................................................................................391
Running selftests