
Remote control commands
R&S
®
NRX
350
User Manual 1178.5566.02 ─ 09
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:RANGe:CLEVel:
STATe
<value>
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:
ON
is set.
Enables or disables the reduction of the transition range between the measurement
paths, set by
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
OFF
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:RANGe:
CLEVel[:VALue]
<value>
Reduces the transition range between the measurement paths, 0 -> 1 and 1 -> 2, by
the set value, the so-called cross-over level. Thus, you can improve the measurement
accuracy for signals with a high peak-to-average ratio, since the headroom for modula-
tion peaks becomes larger. However, the S/N ratio is reduced at the lower limits of the
transition ranges.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
-20.0 to 0.0
*RST:
0.0
Default unit: dB
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:RANGe[:VALue]
<value>
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:
OFF
is set.
Sets the active measurement path in which the power sensor is measuring.
Configuring sensors