Remote control commands
R&S
®
NRX
387
User Manual 1178.5566.02 ─ 09
Suffix:
<output>
.
1 to 2
BNC connectors at the rear; 1
=
Out
1
/
Trig
Out,
2
=
Trig
In
/
Out
2
Parameters:
<value>
Range:
-180.0 to +180.0
*RST:
+10.0
Default unit: dB
Manual operation:
See
OUTPut:TRIGger:SOURce
<source>
TOUT
is set.
Sets the trigger source.
Parameters:
<source>
SENS1 | SENS2 | SENS3 | SENS4 | EXTernal | CHKSource
*RST:
EXTernal
Manual operation:
See
"Trigger Source for Trigger Output"
13.11
Zeroing
Further Information:
●
........................................................................................... 387
..................................................................................388
CALibration<undef>:ALL:ZERO:AUTO
............................................................................. 388
CALibration<Sensor>:ZERO:FAST:AUTO
.........................................................................389
CALibration<undef>:ALL:ZERO:FAST:AUTO
.................................................................... 389
CALibration<Sensor>:ZERO
[<auto>]
Performs zeroing for the power sensor connected to selected port.
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
While zero calibration is in progress, no queries or other setting commands are
allowed, since the command is synchronous. Any communication attempt can run into
a timeout. Use
to recognize the end of a zeroing procedure.
Zeroing