
Remote control commands
R&S
®
NRX
275
User Manual 1178.5566.02 ─ 09
RPLeft
Starting from the left border, at a power difference of the marker
position to the y-position of the reference marker.
RPRight
Starting from the right border, at a power difference of the
marker position to the y-position of the reference marker.
PSEarch
Measured maximum power
MSEarch
Measured minimum power
RPSLeft
Maximum power measured left from the reference marker.
RPSRight
Maximum power measured right from the reference marker.
RMSLeft
Minimum power measured left from the reference marker.
RMSRight
Minimum power measured right from the reference marker.
*RST:
FTIMe
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MARKer<Marker>:POSition:POWer:DBM
<power>
Sets an absolute power value for the marker position defined under
WINDow<Window>]:TRACe:MARKer<Marker>:POSition:MODE
Suffix:
<Window>
.
1 to 4
Measurement pane
<Marker>
1 to 4
Measurement pane
Parameters:
<power>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dBm
Manual operation:
See
DISPlay[:WINDow<Window>]:TRACe:MARKer<Marker>:POSition:POWer:DBUV
<power>
Sets an absolute power value for the marker position defined under
WINDow<Window>]:TRACe:MARKer<Marker>:POSition:MODE
Measurement settings and results