Remote control commands
R&S
®
NRX
453
User Manual 1178.5566.02 ─ 09
[SENSe<Sensor>:]STATistics:SCALe:X:RLEVel
<rlev>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>:STATistics[:SCALe]:X:RLEVel:RELative
CALCulate<Measurement>:STATistics[:SCALe]:X:RLEVel[:ABSolute]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<rlev>
Range:
-280.0 to 220.0
*RST:
-30.0
Default unit: dBm
[SENSe<Sensor>:]STATistics:TIME
<time>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG]:TIME
if a gate is set.
CALCulate<Measurement>:STATistics:APERture
if no gate is set.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
10.0e-6 to 0.3
*RST:
0.01
Default unit: s
[SENSe<Sensor>:]STATistics[:EXCLude]:MID:OFFSet[:TIME]
<time>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:
OFFSet[:TIME]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.0 to 0.3
*RST:
0.0
Default unit: s
Remote emulation