![Rohde & Schwarz 1424.7805K02 Скачать руководство пользователя страница 376](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791376.webp)
Remote control commands
R&S
®
NRX
376
User Manual 1178.5566.02 ─ 09
1
200
kHz
2
Maximum bandwidth of the power sensor
Range:
0 to 2
*RST:
2
Manual operation:
See
[SENSe<Sensor>:]BWIDth:VIDeo:FNUMber
<fnum>
[SENSe<Sensor>:]BANDwidth:VIDeo:FNUMber
<fnum>
Requires the interface for R&S NRT-Z sensors (R&S NRX-B9).
Sets the video bandwidth for the rectified RF. The setting mainly influences the mea-
surement of the peak envelope power (PEP), determination of the crest factor (CF), the
measurement of the average burst power and the complementary cumulative distribu-
tion function (CCDF).
Suffix:
<Sensor>
.
101
R&S NRT directional power sensor
Parameters:
<numeric_value>
0
4
kHz
1
200
kHz
2
The full bandwidth depends on the power sensor:
R&S
NRT
‑
Z14: 600
kHz
R&S
NRT
‑
Z43, R&S
NRT
‑
Z44: 4
MHz
Range:
0 to 2
*RST:
0
Manual operation:
See
13.9
Configuring the test generator
If the sensor check source (R&S NRX-B1) is installed, you can use it as a power refer-
ence for testing the connected power sensors.
Further Information:
●
●
.................................................................................................377
.................................................................................................377
Configuring the test generator