![Rohde & Schwarz 1424.7805K02 Скачать руководство пользователя страница 239](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791239.webp)
Remote control commands
R&S
®
NRX
239
User Manual 1178.5566.02 ─ 09
CALCulate<Measurement>[:CHANnel<Channel>]:FEED<Channel>
"<string>"
Determines the data that are processed. The parameters depend on the measurement
type.
The power sensor averages every measured value using a series of samples. If a
RANDom
feed is selected, the power sensor takes a random value from the samples
and forwards it to the R&S NRX as a measured value.
PEAK
is the maximum of all
samples in the measurement interval.
Table 13-3: Parameter to measurement assignment
Measurement type
"<string>"
Measured value
Continuous average
POWer:AVERage
POWer:PEAK
POWer:RANDom
Average value
Peak value
Randomly selected value
Burst average
POWer:AVERage
POWer:PEAK
POWer:RANDom
Average value
Peak value
Randomly selected value
Trace
POWer:TRACe
POWer:PEAK:TRACe
POWer:RANDom:TRACe
Measurement sequence
Peak value of the samples per test point
Randomly selected value
Pulse analysis
POWer:TRACe
POWer:PEAK:TRACe
POWer:RANDom:TRACe
Measurement sequence
Peak value of the samples per test point
Randomly selected value
Time gate
POWer:AVERage
POWer:PEAK
POWer:RANDom
Average value
Peak value
Randomly selected value
Time gate
Graphical display
The suffix
<1 to
4>
selects the time
gate.
POWer:AVERage ON SWEep<1 to
4>
POWer:PTAVerage ON SWEep<1 to
4>
POWer:PEAK ON SWEep<1 to 4>
Average value in time gate 1, 2, 3 or 4
Peak-to-average value in time gate 1, 2,
3 or 4
Peak value in time gate 1, 2, 3 or 4
Timeslot
POWer:AVERage
POWer:PEAK
POWer:RANDom
Average value
Peak value
Randomly selected value
Statistics
CCDF:TRACe
CDF:TRACe
PDF:TRACe
Complementary cumulative distribution
function
Cumulative distribution function
Probability density function
Measurement settings and results