Measurement and display configuration
R&S
®
NRX
66
User Manual 1178.5566.02 ─ 09
Trigger Mode
Controls the trigger execution depending on the settings under
"Normal"
Continuous triggering with regular trigger events.
"Freerun"
Enables a continuous measurement. The power sensor executes one
measurement cycle after the other.
"Single"
Disables continuous triggering so that only one trigger event at a time
is executed. To enable triggering again, press [1Trig].
"Auto"
Automatically starts a measurement if no trigger event has occurred
after 300
ms.
Remote command:
TRIGger<undef>:ALL:MODE
TRIGger<Measurement>:MODE
Trigger Source
For primary and secondary power sensors.
Sets the source for the trigger event. See
Chapter 6.1.3.2, "Trigger sources"
Remote command:
TRIGger<undef>:ALL:SOURce
TRIGger<Measurement>[:CHANnel<Channel>]:SOURce
Trigger Level
For primary and secondary power sensors.
Sets the trigger threshold for internal triggering derived from the test signal.
The trigger level is displayed as dotted red line. The trigger time is displayed as small
rhomb on the trigger level line.
Remote command:
TRIGger<undef>:ALL:LEVel
TRIGger<Measurement>[:CHANnel<Channel>]:LEVel
Trigger Advanced
For primary and secondary power sensors.
Groups further trigger settings.
Source ← Trigger Advanced
See
Slope ← Trigger Advanced
Determines which edge of the envelope power (internal triggering) or increasing volt-
age (external triggering) is used for triggering.
"Positive"
Rising edge
"Negative"
Falling edge
Configuration for all measurement types