Remote control commands
R&S
®
NRX
454
User Manual 1178.5566.02 ─ 09
[SENSe<Sensor>:]STATistics[:EXCLude]:MID:TIME
<time>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:POWer]:TGATe<Gate>[:AVG][:EXCLude]:MID:
TIME
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<time>
Range:
0.0 to 0.3
*RST:
0.0
Default unit: s
[SENSe<Sensor>:]TIMing:EXCLude:STARt
<exclude_start>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>:TSLot:TIMing:EXCLude:STARt
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<exclude_start>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
[SENSe<Sensor>:]TIMing:EXCLude:STOP
<exclude_stop>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>:TSLot:TIMing:EXCLude:STOP
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<exclude_stop>
Range:
0.0 to 15.0
*RST:
0.0
Default unit: s
[SENSe<Sensor>:]TRACe:AVERage:COUNt
<filter length>
Available for compatibility. Recommended R&S NRX command:
Remote emulation