Remote control commands
R&S
®
NRX
309
User Manual 1178.5566.02 ─ 09
13.6.12
Statistics measurements
There are several ways to programm the R&S NRX for measurements. See
ter 13.4, "Making measurements"
Further information:
●
13.6.12.1
Combining commands for statistics measurements
See also
Chapter 13.4, "Making measurements"
Parameter list for the statistics measurement
Parameter
Description
<statistics_size>
Mandatory. Number of test points on the time axis.
Corresponds to
[SENSe<Sensor>:]STATistics:SCALe:X:POINts
<capture_time>
Mandatory. Period within which measured data are captured in the trace
measurements.
Corresponds to
<source_list>
Mandatory. Primary and secondary sensor. The number for each sensor
is preceded by the character
@
. The entire expression is enclosed in
parentheses.
Example: (
@
3),(
@
2)
Sensor
C is the primary sensor, and sensor
B is the secondary sensor.
CONFigure<Measurement>:STATistics:CCDF?
<statistics_size>, <capture_time>,
<source_list>
FETCh<Measurement>:STATistics:CCDF?
<statistics_size>, <capture_time>,
<source_list>
READ<Measurement>:STATistics:CCDF?
<statistics_size>, <capture_time>,
<source_list>
MEASure<Measurement>:STATistics:CCDF?
<statistics_size>, <capture_time>,
<source_list>
Measures the power in the defined time interval and performs a statistic evaluation
(probability density function, PDF).
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<statistics_size>
<expr>
<capture_time>
Default unit: s
Measurement settings and results