Remote control commands
R&S
®
NRX
217
User Manual 1178.5566.02 ─ 09
13.6.1.3
Units
If you enter a value that is expressed in a certain unit, for example Hz, you can omit
the unit. Then, the default unit provided in the remote command description is used. If
you need decimal multiples and submultiples of a unit, you can use SCPI prefixes, see
If you enter a power value or power ratio that can be expressed in more than one unit,
you can enter the value together with the unit, and the unit is recognized. If you enter a
value without unit, the unit defined by one of the following commands is used:
●
UNIT<Measurement>:POWer[:VALue]
●
After a reset, the default unit is used.
SENSe<Sensor>:UNIT:POWer[:VALue]
............................................................................217
................................................................................ 218
UNIT<Measurement>:POWer:REFLection
........................................................................218
UNIT<Measurement>:POWer:RELative:STATe
................................................................. 218
UNIT<Measurement>:POWer[:VALue]
............................................................................. 219
SENSe<Sensor>:UNIT:POWer[:VALue]
<unit>
Sets the unit for power-related
SENSe
parameters if you enter a value without unit.
Information whether this command applies is given in the decription of the affected
command.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<unit>
DBM | DBUV | W
Available units.
If the R&S
NRP2 emulation is enabled using
*RST:
DBM
SOURce:UNIT:POWer
<power>
Requires the sensor check source (R&S NRX-B1).
Sets the unit of the power level for the output signal.
Parameters:
<power>
DBM | DBUV | W
*RST:
DBM
Manual operation:
See
Measurement settings and results