Measurement and display configuration
R&S
®
NRX
86
User Manual 1178.5566.02 ─ 09
Start Time ← Time tab
Time when the signal passes through the medial reference power level with rising
edge, referenced to the delayed trigger event. Indicates the start point of the first power
pulse within the analysis window.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:POSitive:
OCCurrence[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:POSitive:
OCCurrence?
Stop Time ← Time tab
Time when the signal passes through the medial reference power level with falling
edge, referenced to the delayed trigger event. Indicates the stop point of the first power
pulse within the analysis window.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:TRANsition:NEGative:
OCCurrence[:STATe]
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:NEGative:
OCCurrence?
Sampling Rate ← Time tab
Number of samples per second.
Remote command:
DISPlay[:WINDow<Window>]:TRACe:MEASurement:PULSe:RESolution[:
STATe]
CALCulate<Measurement>:TRACe:MEASurement:TRANsition:SPERiod?
Power tab
Selects the power-related measurement results that are displayed below the trace. The
selections on this tab apply to all traces.
Pulse analysis