Remote control commands
R&S
®
NRX
342
User Manual 1178.5566.02 ─ 09
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:MTIMe
<value>
Effective for continuous average, burst average measurements, if
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
TYPE
RES
is set.
Sets an upper limit for the settling time of the auto-averaging filter that is never excee-
ded, thus limiting the length of the filter.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
0.01 to 1000.0
*RST:
1.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
NSRatio
<value>
Effective for continuous average, burst average measurements, if
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:
TYPE
NSR
is set.
Sets the averaging number so that the intrinsic noise of the power sensor does not
exceed the specified value.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
Range:
100e-6 to 1.0
*RST:
1.0
Default unit: dB
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO:SLOT
<value>
Effective for continuous average, burst average measurements.