Remote control commands
R&S
®
NRX
270
User Manual 1178.5566.02 ─ 09
<capture_time>
Default unit: s
<source_list>
<expr>
For the parameter descriptions, see
Usage:
Query only
CONFigure<Measurement>:XTIMe[:POWer]:NONE
<scope_size>, <capture_time>,
<source_list>
Disables trace
2. In contrast, trace
1 is always active.
Suffix:
<Measurement>
.
1 to 8
Measurement
Setting parameters:
<scope_size>
<expr>
<capture_time>
Default unit: s
<source_list>
<expr>
For the parameter descriptions, see
Usage:
Setting only
CONFigure<Measurement>:XTIMe[:POWer]:RATio?
<scope_size>, <capture_time>,
<source_list>
FETCh<Measurement>:XTIMe[:POWer]:RATio?
<scope_size>, <capture_time>,
<source_list>
READ<Measurement>:XTIMe[:POWer]:RATio?
<scope_size>, <capture_time>,
<source_list>
MEASure<Measurement>:XTIMe[:POWer]:RATio?
<scope_size>, <capture_time>,
<source_list>
Power ratio over time measured by two power sensors.
Suffix:
<Measurement>
.
1 to 8
Measurement
Query parameters:
<scope_size>
<expr>
<capture_time>
Default unit: s
<source_list>
<expr>
For the parameter descriptions, see
Usage:
Query only
Measurement settings and results