Measurement and display configuration
R&S
®
NRX
65
User Manual 1178.5566.02 ─ 09
If you use a hold-off time instead of a dropout time, you can obtain stable triggering
conditions - regular triggering at the same point. But you cannot achieve exclusive trig-
gering at A.
6.1.3.4
Hold-off time
During the hold-off time, a period after a trigger event, all trigger events are ignored.
6.1.3.5
Trigger settings
Access: Main measurement dialog
> "Trigger"
For trace or pulse analysis measurements, the access is: Main measurement dialog
>
"Trigger"
> "Sensor Trigger"