
Remote control commands
R&S
®
NRX
242
User Manual 1178.5566.02 ─ 09
RNDMax
Randomly selected samples are transmitted. All evaluations use
these values instead of the average values.
*RST:
NONE
13.6.4
Selecting the power sensor
Further information:
●
Chapter 6.1.4, "Measurement settings dialog"
........................................................................................... 242
CALCulate<Measurement>[:CHANnel<Channel>]:SENSe:INDex
........................................242
[SENSe<Sensor>:]CATalog?
Returns a list of all connected power sensors together with the suffix of the port where
the power sensor is connected.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Example:
CAT?
Query
"2:NRP33SN-900004"
Response
Usage:
Query only
Manual operation:
See
"Primary Sensor, Secondary Sensor"
CALCulate<Measurement>[:CHANnel<Channel>]:SENSe:INDex
<index>
Effective for continuous average, burst average, trace, pulse analysis, time gate, time-
slot measurements.
Assigns the primary or secondary sensor. You can choose any of the sensors that are
connected to a sensor port of the R&S NRX. You can query the connected power sen-
sors using
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<index>
Selects the power sensor by the R&S NRX port to which it is
connected.
Measurement settings and results