Remote control commands
R&S
®
NRX
443
User Manual 1178.5566.02 ─ 09
Usage:
Query only
[SENSe<Sensor>:]INPut:ATTenuation:AUTO
<auto>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:INPut:ATTenuation:AUTO
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<auto>
OFF | ON | ONCE
*RST:
OFF
[SENSe<Sensor>:]INPut:ATTenuation[:VALue]
<value>
Available for compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:INPut:ATTenuation[:
VALue]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
Range:
0.0 to 30.0
*RST:
30.0
Default unit: dB
[SENSe<Sensor>:]INTernal:TRIGger:JITTer:METHod
<method>
Available for compatibility. Recommended R&S NRX command:
TRIGger<Measurement>[:CHANnel<Channel>]:JITTer:METHod
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<method>
COMPensate | MEASure | NONE
*RST:
COMPensate
[SENSe<Sensor>:][POWer:][AVG:]APERture[:VALue]
<integration_time>
Available for compatibility. Recommended R&S NRX command:
Remote emulation