![Rohde & Schwarz 1424.7805K02 Скачать руководство пользователя страница 407](http://html1.mh-extra.com/html/rohde-and-schwarz/1424-7805k02/1424-7805k02_user-manual_1477791407.webp)
Remote control commands
R&S
®
NRX
407
User Manual 1178.5566.02 ─ 09
Usage:
Query only
SYSTem:HELP:SYNTax:ALL?
Queries the implemented SCPI commands and their parameters. Returns the result as
a block data.
Return values:
<Syntax>
<dblock>
Usage:
Query only
SYSTem:IDN:ANSWer
<string>
USER
is set.
Sets the customized instrument identification string so that you can identify each R&S
NRX individually.
You can query the defined string using
Parameters:
<string>
Identification string. Maximum string length is 128 characters.
Manual operation:
See
SYSTem:IDN:AUTO
<status>
Enables or disables the automatic instrument identification for
Parameters:
<status>
ON | OFF
*RST:
1
Manual operation:
See
SYSTem:IDN:MODE
<mode>
Sets which identification string is used.
Parameters:
<mode>
AUTO | USER
AUTO
Automatic instrument identification.
USER
Customized identification string. Define the string using
*RST:
AUTO
Manual operation:
See
System information and configuration