Remote control commands
R&S
®
NRX
308
User Manual 1178.5566.02 ─ 09
Parameters:
<value>
Range:
50.0e-9 to 0.1
*RST:
1.0e-3
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID:OFFSet[:TIME]
<value>
Sets the length of the fence.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.0 to 0.1
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID:TIME
<value>
Sets the start time of the fence. The start time refers to the start of the timeslots.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
0.0 to 0.1
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID[:STATe]
<value>
Enables or disables an exclusion interval in the timeslots. This exclusion interval is
called fence. The fence is defined by its start time and its length.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
*RST:
OFF
Manual operation:
See
Measurement settings and results