Remote control commands
R&S
®
NRX
236
User Manual 1178.5566.02 ─ 09
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<port>
INTernal | EXT2 | EXTernal2
*RST:
INTernal
Manual operation:
See
TRIGger<Measurement>[:CHANnel<Channel>]:SENDer[:STATe]
<state>
Enables or disables the power sensor as trigger sender. If enabled, the power sensor
outputs a digital trigger signal in sync with its own trigger event. The trigger signal is
output at the port selected by
TRIGger<Measurement>[:CHANnel<Channel>]:
.
The trigger sender has to use its internal trigger source. Set the trigger source for the
trigger receivers to
INTA
|
INTB
|
INTC
|
INTD
, where [A to D] is the port to which the
trigger sender is connected. The trigger signal generated by the trigger sender is
routed to the R&S NRX and from there it is distributed to the trigger receivers and, if
SENS1
|
SENS2
|
SENS3
|
SENS4
is set, also to the trigger
output.
The command is not fully R&S
NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S
NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<state>
ON | OFF
*RST:
OFF
Manual operation:
See
TRIGger<undef>:ALL:SLOPe
<slope>
TRIGger<Measurement>[:CHANnel<Channel>]:SLOPe
<slope>
Determines which edge of the envelope power (internal triggering) or increasing volt-
age (external triggering) is used for triggering.
The command is not fully R&S
NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S
NRP2 emulation using
"NRP2"
.
Measurement settings and results