Remote control commands
R&S
®
NRX
345
User Manual 1178.5566.02 ─ 09
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:TYPE
<type>
Requires an R&S
frequency selective power sensor.
Sets the averaging method. See also the user manual of the power sensor.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<type>
POWer | VIDeo | LINear
*RST:
POWer
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage[:STATe]
<value>
Enables or disables the averaging filter. If enabled, the number of measured values is
averaged. Averaging reduces the effect of noise so that more reliable results are
obtained.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage:COUNt[:
VALue]
<value>
Effective for trace measurements.
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Suffix:
<Measurement>
.
1 to 8
Measurement
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Configuring sensors